Recall Z-2773-2026 ยท OUT Intracranial Aneurysm Flow Diverter

Micro Therapeutics, Inc.

Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12. Intracranial aneurysm flow diverter

DateJul 22, 2026 (initiated Jun 8, 2026)
StatusOpen, Classified
ReasonIncorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.
Root causeLabeling Change Control
ActionOn June 4, 2026, Medtronic (parent company of Micro Therapeutics) issued a Urgent Medical Device Expansion to affected consignees via visit. Medtronic ask consignees to take the following actions: 1. Do NOT use any impacted Pipeline Shield products. Remove and quarantine all unused impacted product units in your inventory. 2. Return the impacted product(s) to Medtronic. Your Medtronic representative can assist in facilitating the return of product and/or identifying suitable replacement product as necessary. 3. Complete the attached Customer Confirmation Form and email it to neuro.quality@m
Quantity11 units
DistributionUS Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.