Recall Z-2801-2026 ยท OJG Neurological Tray

Windstone Medical Packaging, Inc.

AMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: 1) Neuro Pack, Pack Number AMS12195C; 2) Neuro Pack, Pack Number AMS12195D; 3) Neuro Pack, Pack Number AMS12307; 4) Neuro Spine Pack, Pack Number AMS16313; 5) Lami Kukkar Pack, Pack Number AMS17958; 6) Neuro Pack, Pack Number AMS6452B; 7) Neuro Lami Pack, Pack Number AMS9427C.

DateJul 24, 2026 (initiated Jun 8, 2026)
StatusOpen, Classified
ReasonThe recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
Root causeMaterial/Component Contamination
ActionAligned Medical Solutions (AMS) issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 6/8/2026 via email. The notice explained the issue, potential risk, and requested the following actions: identify and quarantine affected product, affix the sticker which reads: "RECALL NOTICE Medicom announced the recall of Ritmed Neurological Sponges contained in this kit after it was sealed and sterilized. All other components in the kit are not affected by this recall." " At the time the kit is opened for use any Medicom-Ritmed Neurological Sponges should be identified and set aside.
Quantity927 units
DistributionUS Nationwide distribution in the states of AZ, CA, CT, GA, IL, MA, MD, NY, OH, PA, TN, TX, VA, WA.