Recall Z-2954-2026 ยท EFD Amalgamator, Dental, Ac-Powered

SDI Limited

Dentsply Sirona Capsule Mixer Model No. 5546068

DateAug 14, 2026 (initiated Jul 20, 2026)
StatusOpen, Classified
ReasonPotential for electrical discharge from unit when unplugged.
Root causeDevice Design
ActionOn July 20, 2026, Dentsply Sirona issued a Urgent Medical Device Recall notice to affected consignees via mail. Dentsply Sirona ask consignees to take the following actions: Users 1. Switch off the unit using the rear power switch before removing the plug from the main socket 2. Avoid immediate contact with the plug pins after unplugging. 3. The users should return their unit to Dentsply Sirona to have a corrective modification installed. Please contact Racheal Elliott at rachael.elliott@dentsplysirona.com. 4. Please ensure that all relevant personnel within your organization are informed
Quantity1,444 units
DistributionUS: AK AL AR AZ CA CO CT DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV OUS: None