ClearanceBrief
Recall Z-2811-2026 · DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Medtronic Perfusion Systems

EOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78322

DateJul 27, 2026 (initiated Jun 24, 2026)
StatusOpen, Classified
ReasonPinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Root causeUnder Investigation by firm
ActionAn URGENT: MEDICAL DEVICE RECALL notification letter dated June 2026 was sent to customers beginning 6/24/26. Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for listed products using Attachment A. " Immediately identify and quarantine all unused, listed product in your inventory. " Return unused, listed product in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return for credit or replacement product. Your Medtronic sales representative can a
Quantity1,424 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K260195Elongated One-Piece Arterial (EOPA) 3D™ Arterial Cannulae
Medtronic, Inc., Apr 23, 2026