Recall Z-2809-2026 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Medtronic Perfusion Systems
EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77722
| Date | Jul 27, 2026 (initiated Jun 24, 2026) |
| Status | Open, Classified |
| Reason | Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction. |
| Root cause | Under Investigation by firm |
| Action | An URGENT: MEDICAL DEVICE RECALL notification letter dated June 2026 was sent to customers beginning 6/24/26. Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for listed products using Attachment A. " Immediately identify and quarantine all unused, listed product in your inventory. " Return unused, listed product in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return for credit or replacement product. Your Medtronic sales representative can a |
| Quantity | 3,110 units |
| Distribution | Worldwide - US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K031037 | EOPA ELONGATED ONE-PIECE ARTERIAL CANNULA AND GUIDEWIRE Medtronic Perfusion Systems, May 13, 2003 |