ClearanceBrief
Recall Z-2786-2026 ยท OHD Gynecological Laparoscopic Kit

Medline Industries, LP

Medline convenience kits containing Bard Silastic Foley Catheters (French sizes 16, 18, 20) 5cc balloons. SM-LAPAROSCOPIC PROSTATE PAC DYNJ61206J

DateJul 23, 2026 (initiated Jun 8, 2026)
StatusOpen, Classified
ReasonAffected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
Root causeNonconforming Material/Component
ActionOn June 8, 2026, the firm began notifying customers through First-Class Mail and email. Customers were instructed to over-label affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity980 kits
DistributionUS Nationwide distribution in the states of CA, FL, MD. PA, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.