ClearanceBrief
Recall Z-2852-2026 ยท ITI Wheelchair, Powered

Pride Mobility Products Corp.

Brand Name: Quantum Rehab Product Name: 4Front Model/Catalog Number: 4Front Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller

DateJul 31, 2026 (initiated Jun 25, 2026)
StatusOpen, Classified
ReasonIf the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
Root causeUnder Investigation by firm
ActionFirm notified dealers and consumers via mailed letters on June 25, 2026. Dealers were instructed to locate affected units and inform consumers of the required software update. Dealers should schedule service appointments with affected consumers. Systems should be updated to Q-Logic software version 2.0.13 or later. The update must be completed following a system power cycle. Consumers were instructed to disable the "Remind Me Later" feature immediately. Consumers should contact their dealer to schedule an appointment for software update.
Quantity1102
DistributionWorldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Austra
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143383Q6 Edge EM
Pride Mobility Product Corporation, Jul 23, 2015