Recall Z-2760-2026 ยท FPO Stretcher, Wheeled

Ferno-Washington Inc

Power F2 Powered Fastening System, Product No. 0567091 Powered fastening system intended for professional use only. Designed to load, unload, and secure a compatible wheeled stretcher inside a ground-based ambulance. Item no. 0567091

DateJul 17, 2026 (initiated Jun 19, 2026)
StatusOpen, Classified
ReasonProduct may experience accelerated roller wear and potential deformation due to interference between the roller assembly and rail. In the worst-case condition, rail mechanism may become lodged during operation and movement of the stretcher along the rail may no longer be possible through either powered or manual operation. May cause delays in transport of patient to healthcare facility.
Root causeProcess control
ActionOn June 19. 2026, the firm began notifying customers via Urgent Medical Device Correction letters. Customers were instructed to contact Ferno for assistance if abnormal rail movement, increased resistance during rail extension or retraction, or other unusual rail operation is observed. Ferno will contact all affected customers directly and dispatch trained field service personnel to perform the correction onsite at customer locations. The correction does not require removal of the PF2 Fastening System from service prior to the scheduled repair. The field correction consists of replacement o
Quantity278
DistributionWorldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, GA, IA, ID, IL, IN, KY, LA, MD, ME, MN, MO, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Poland, Netherlands, Canada, Germany, Norway, United Kingdom, Denmark and New Zealand.