Recall Z-2761-2026 ยท GCJ Laparoscope, General & Plastic Surgery

Olympus Corporation of the Americas

Olympus 4K UHD LCD Monitor. Model Number: OEV321UH. An electrically powered visual display unit designed for displaying graphical information in color.

DateJul 20, 2026 (initiated Jul 7, 2026)
StatusOpen, Classified
ReasonPotential for the HDMI and DisplayPort (DP) connectors can become faulty resulting in image loss or image abnormalities.
Root causeNonconforming Material/Component
ActionOn July 7, 2026, the firm issued "Urgent: Medical Device Correction" letters to consignees. Customers were informed of the product issue. While the identified 4K UHD LCD monitor may be used with VISERA S, VISERA ELITE III, EVIS X1, and EU-ME3 systems, the firm notes that the affected HDMI or DP connectors are more commonly used with the VISERA S video system center. Its use with the three other systems does not typically involve the affected HDMI or DP connectors. Customers are requested to examine their equipment, looking for the identified monitor model number and serial numbers listed i
Quantity10 units
DistributionWorldwide - US Nationwide distribution in the states of OH, CA, NJ, FL and the country of Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K2200694K UHD LCD Monitor OEV321UH
Olympus Medical Systems Corp., Mar 9, 2022