Recall Z-2805-2026 ยท ITI Wheelchair, Powered

Permobil Inc

M5 Corpus Power Wheelchair REF,120032

DateJul 24, 2026 (initiated Jun 19, 2025)
StatusOpen, Classified
ReasonThis recall is being reported retroactively. Power Platform firmware is missing specific drive inhibit logic. This issue may allow the chair to drive under restricted conditions, and may create an unstable condition in certain circumstances.
Root causeProcess design
ActionOn June 10, 2025 Permobil issued a technical bulletin (recall notice) via Email to affected consignees. Permobil informed consignees of the following: 1. Cellular connection is required to receive and install the firmware update. 2. It is important to not use the wheelchair when the firmware update is in process (as indicated on the display of the input device). 3. Specific Remote Update Instructions are available if needed in addition to the on-screen information. 4. A confirmation of the completed update will be shown on the joystick display when the update successfully completes.
Quantity715 wheelchairs
DistributionWorldwide - US: Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CANADA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123290M300 & M400
Permobil AB, Apr 17, 2013