Recall Z-2937-2026 · PIG Acute Kidney Injury Test System

Bioporto Diagnostics A/S

ProNephro AKI (NGAL) Reagent Kit (100 tests). IVD Acute Kidney Injury Test System.

DateAug 12, 2026 (initiated May 18, 2026)
StatusOpen, Classified
ReasonThe use of a non-validated rework process for assay reagents which could lead to missed diagnosis of renal disease, inappropriate treatment decisions, or inappropriate monitoring of renal disease in intended users.
Root causeNonconforming Material/Component
ActionA Quality Notification dated 6/3/26 was sent to consignees via email notifying them of the recall. Consignees in the US are to coordinate replacement of devices. ROW Consignees are to scrap unopened kits in accordance with local requirements and request replacements.
Quantity400 units
DistributionWorldwide - The affected product distribution included the United States and the countries of European Union, Chile, and Hong Kong. No affected product was distributed to Canada; however, Health Canada was notified because the product is registered in Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232761ProNephro AKI™ (NGAL)
Bioporto Diagnostic, Inc., Dec 7, 2023