Recall Z-2941-2026 ยท FPA Set, Administration, Intravascular

BD SWITZERLAND SARL

60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus Ext Set w/Needleless and Y-Site REF MPX5305-C UDI-DI code: 10885403274619 9 IN Extension Set w/Max Y REF MPX5300-C UDI-DI code: 10885403237218 The Intravascular Administration Set and Extension Set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter

DateAug 13, 2026 (initiated Jun 16, 2026)
StatusOpen, Classified
ReasonDue to mechanical issue with internal sterilization equipment resulting in potential underdosing of product during processing, which may have resulted in products not being sterilized correctly.
Root causeProcess control
ActionOn 06/16/2026, an "URGENT: Medical Device Recall" Letter was emailed to customers informing them that, BD has identified a mechanical issue with internal sterilization equipment resulting in potential underdosing of product during processing, which may have resulted in product not being sterilized correctly. Customer are instructed to: "Immediately discontinue use of the affected lot(s). "Check all inventory locations within your facility and destroy all affected units in accordance with your facility s process. If destruction is not immediately possible at your facility, then product should
Quantity93,700
DistributionU.S.: AZ, CA, CO, FL, GA, IL, IN, KS, MD, MI, MN, MO, MS, NC, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA O.U.S.: Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051499INTRAVASCULAR ADMINISTRATION SET AND EXTENSION SET
Medegen Medical Manufacturing Services, Jun 22, 2005
K083472MEDEGEN PRESSURE RATED EXTENSION SETS
Medegen, Inc., Dec 9, 2008