Recall Z-2772-2026 ยท OES Cardiac Catheterization Kit

AVID Medical, Inc.

Halyard STANDARD LINE INSERTION PACK. Model/Catalog Number: PVHM234-01. Convenience kit.

DateJul 22, 2026 (initiated May 19, 2026)
StatusOpen, Classified
ReasonDefects in the silicone seal, which may result in "doming" of the silicone seal due to improper sealing, potentially leading to fluid occlusion or leakage.
Root causeProcess change control
ActionAVID Medical notified consignees on about 05/19/2026 via email. Consignees were instructed to identify, segregate, and quarantine all impacted kits, and add the provided warning labels to affected AVID kits to ensure all users are aware of discarding the impacted component prior to use. Distributors were instructed to ensure all end-users are properly notified if affected kits have been further distributed. Customers/End-users were instructed that if they have received product from a distributor, to contact the distributor or your sales representative with any questions. Recipients were reques
Quantity112 kits
DistributionU.S. Nationwide distribution in the state of California.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.