Recall Z-2832-2026 ยท KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component

Stryker GmbH

Stryker Hoffmann II Connecting Rod Micro 3 x 120mm. Catalog Number: 50796120.

DateJul 29, 2026 (initiated Jul 1, 2026)
StatusOpen, Classified
ReasonPotential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
Root causeDevice Design
ActionFirm began notifying consignees on July 1, 2026. Customers were instructed to immediately check their internal inventory to locate affected product. Customers were informed that until further notice, the subject devices are limited to single use only. Following first use, discard the devices in accordance with the facility's established procedures. Users should conduct visual inspection of product prior to each use. If cracking is found to be present, the customer should quarantine non-confirming parts and report the issue to Stryker's Trauma & Extremities division. Rods with cracks shoul
Quantity5,439 units
DistributionWorldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, ID, IL, KS, LA, MA, MD, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, PA, RI, SC, TN, TX, VA, WI, WV and the countries of Australia, Austria, Belgium, Canada, Chile, Denmark, France, Germany, Greece, Ireland, I
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K033476MODIFICATION TO HOFFMANN II MICRO EXTERNAL FIXATION SYSTEM
Howmedica Osteonics Corp., Jan 20, 2004