Recall Z-2762-2026 ยท NKB Thoracolumbosacral Pedicle Screw System

SEASPINE ORTHOPEDICS CORPORATION

SeaSpine Polyaxial Head REF Number: 41-3010

DateJul 20, 2026 (initiated May 28, 2026)
StatusOpen, Classified
ReasonDue to manufacturing error, polyaxial head may popoff from the screw intraoperatively following insertion.
Root causeProcess change control
ActionOn June 16, 2026, Orthofix issued a updated Urgent Medical Device Recall notification to provide additional information concerning the recall. Initial notification was issued on May 28, 206. Orthofix ask consignees to take the following actions: 1. Discontinue use of affected lots, 2.Immediately identify and quarantine any affected product. 3. If product is identified, Orthofix will facilitate the return. 4. Complete and return the attached Acknowledgement and Receipt Form. 5. This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to an
Quantity796 units
DistributionWorldwide - US Nationwide distribution including in the states of Alabama, Arizona, California, Colorado, Connecticut, Florida, Illinois, Indiana, Kansas, Maryland, Michigan, Missouri, Mississippi, Montana, Nevada, Ohio, Oklahoma, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173882SeaSpine Mariner Pedicle Screw System
SeaSpine Orthopedics Corporation, Feb 6, 2018