Recall Z-2715-2026 ยท LON System, Test, Automated, Antimicrobial Susceptibility, Short Incubation

Biomerieux Inc

VITEK 2 AST-N447 TEST KIT Product reference 424620

DateJul 9, 2026 (initiated Jun 4, 2026)
StatusOpen, Classified
ReasonPotential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.
Root causeProcess change control
ActionFirm began notifying customers on June 4, 2026. Customers are instructed to dispose of pouches from the affected lot (Ref 424620, Lot 0853293604). Customer may choose to inspect each pouch in their inventory and only dispose of those with "P7" on the label, or they may choose to dispose of all cards from Lot 0853293604 in their inventory. Based on individual customer policy, customers may wish to perform retrospective review of resistant results. Customers should distribute the information to all appropriate personnel in their laboratory. If product was transferred or further distributed, c
Quantity45,900 cards
DistributionInternational distribution to the countries of Belgium and the Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.