Recall Z-2714-2026 ยท GAD Retractor

Integra LifeSciences Corp. (NeuroSciences)

OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

DateJul 9, 2026 (initiated Jun 5, 2026)
StatusOpen, Classified
ReasonThe current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Root causeLabeling design
ActionIntegra issued an URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION notice to its consignees on 06/05/2026 via overnight mail. The notice explained the issue, hazards and risks, and requested the following: "Customer/Medical Facility-Actions to Take: 1. If your facility has validated its own sterilization process, it is in your discretion to follow your validated process recognizing that Integra LifeSciences has only validated the sterilization process noted below. 2. If utilizing the Gravity Sterilization Method, stop using this method immediately. The Gravity Sterilization Method for sterilizin
Quantity633 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Azerbaijan, Bangladesh, Belgium, Brunei Darussalam, Bulgaria, Canada, Chile, China, Czechia, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, J
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.