Recall Z-2647-2026 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayCare 2024A SP4. Oncology Information Systems.

DateJul 1, 2026 (initiated May 20, 2026)
StatusOpen, Classified
ReasonBeam set delivery note issues in which there is potential data loss if the beam set delivery note is modified simultaneously and/or if a note is updated during a treatment session, the user is not notified of the change.
Root causeUnder Investigation by firm
ActionA Field Safety Notice, Medical Device Correction #175428 dated 6/2/26 was emailed to consignees. The notice advises consignees to ensure clear verbal or procedural communication when updating the beam set delivery note, avoid making changes from any location other than the control room when treatment is ongoing, avoid relying on beam set delivery note for conveying critical information. Additionally, consignees are to ensure planning staff and users are educated of the provided workaround. Consignees are to verify the software version of their units and reply to the emailed notification. A sof
Quantity3 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K242992RayCare (2024A SP1)
RaySearch Laboratories AB (PUBL), Apr 17, 2025