Recall Z-2536-2026 ยท OES Cardiac Catheterization Kit

Medline Industries, LP

Medline Convenience Kits: 1) MULTIMED, 7FR, 20CM, 3L, CVC BUNDLE, Model Number: ECVC5995; 2) TRIPLE LUMEN CENTRAL LINE BUNDLE, Model Number: ECVC2625; 3) INTROFLEX 6F, 1L, TB, US, Model Number: STCVC26US; 4) AVA HIGH FLOW, Model Number: ECVC3720; 5) CDP, MM, 3L, 7F, 16CM, H, Model Number: 3K16H1838; 6) CVC Insertion Bundle 3L 7FR 20CM, Model Number: ECVC5610; 7) TRIPLE LUMEN CVC INSERTION BU

DateJun 29, 2026 (initiated May 21, 2026)
StatusOpen, Classified
ReasonThe kits containing Huons Lidocaine HCL Injection USP 1% 10mg/ml 5mL Ampules (Lidocaine Ampules)(OEM). A recall notice was issued by Huons Co., Ltd. for Lidocaine Ampules due to quality issues identified during a recent FDA manufacturing inspection of its drug manufacturing site.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 5/21/26 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-082-FGX2
Quantity967 kits
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall