Recall Z-2914-2026 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

W L Gore & Associates, Inc.

GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VSWVN1010. Flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed.

DateAug 7, 2026 (initiated May 22, 2026)
StatusOpen, Classified
Reason10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010).
Root causeProcess control
ActionOn 05/22/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was sent via email to customers informing them that, 10 mm diameter x 7 cm with transmural drainage holes (VSWVH1006) incorrectly labeled as 10 mm diameter x 10 cm length devices without transmural drainage holes (VSWVN1010). Customers are instructed to: . Identify and return any unused devices within the scope of this recall for replacement. . Complete and sign the enclosed CUSTOMER RESPONSE / DEVICE RECONCILIATION FORM and return to Ashley.Mitch@Olympus.com within 2 weeks of receipt of this notification. . This notice needs to be share
Quantity13 units
DistributionU.S. Nationwide distribution in the states of GA and RI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170740GORE VIABIL Short Wire Biliary Endoprosthesis
W. L. Gore & Associates, Inc., May 8, 2017