Recall Z-2532-2026 ยท HRS Plate, Fixation, Bone

Acumed LLC

Acumed Hexalob Screw various sizes: REF Numbers: : 3070-27008 2.7mm x 8mm NL Hexalobe Screw; 3070-27009 2.7mm x 9mm NL Hexalobe Screw; 3070-27010 2.7mm x 10mm NL Hexalobe Screw; 3070-27011 2.7mm x 11mm NL Hexalobe Screw; 3070-27012 2.7mm x 12mm NL Hexalobe Screw; 3070-27013 2.7mm x 13mm NL Hexalobe Screw; 3070-27014 2.7mm x 14mm NL Hexalobe Screw; 3070-27015 2.7mm x 15mm NL Hexalobe Screw; 3070-27

DateJun 23, 2026 (initiated May 28, 2026)
StatusOpen, Classified
ReasonDue to manufacturing error, hexalobe screws used with wrist fixation system may break during use.
Root causeDevice Design
ActionOn May 28, 2026 Acumed issued a Urgent Medical Device Recall Notification to affected consignees via phone calls and email. Acumed asked consignees to take the following actions: 1)IMMEDIATELY stop using and/or distributing the product and complete the following actions: 2) Perform a physical count of your inventory. Identify any inventory of the batch/lot numbers listed in the table above. 3) Remove the product segregated in step 1, such that they are taken out of service (i.e., quarantined). 3) Verbally acknowledge (then respond via email) to Kathryn.jayne@acumed.net the recall notice.
Quantity5,272 units
DistributionUS Nationwide distribution in the states of AR, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MS, NC, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K251132The Acumed Wrist Fixation System - Plates; The Acumed Wrist Fixation System - Screws
Acumed, LLC, Jul 14, 2025
K251296The Acumed Wrist Fixation System
Acumed, LLC, Jun 18, 2025