ClearanceBrief
Recall Z-2550-2026 ยท NHJ Device, Positive Pressure Breathing, Intermittent

Baxter Healthcare Corporation

Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device

DateJul 1, 2026 (initiated May 21, 2026)
StatusOpen, Classified
ReasonThere have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Root causeNonconforming Material/Component
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECITON notice to its consignees on 5/21/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: "Actions to be Taken by Customers 1. Before each use, carefully inspect the nebulizer cup for damage and ensure it is fully and securely locked during assembly, following the enclosed instructions. If leakage is observed during therapy, discontinue the use of that nebulizer and replace the patient circuit. 2. Nebulizer cups with patient circuits are routinely replaced as part of normal therapy usage
Quantity16187 units
DistributionWorldwide distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192143Maximus System
Hill-Rom Services Pte, Ltd., Feb 14, 2020
K200988Maximus System
Hill-Rom Services Pte, Ltd., May 26, 2020