Recall Z-2776-2026 ยท HIF Insufflator, Laparoscopic

Olympus Corporation of the Americas

Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4

DateJul 22, 2026 (initiated Jun 16, 2026)
StatusOpen, Classified
ReasonPotential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
Root causeNonconforming Material/Component
ActionOlympus notified consignees on about 06/16/2026 via letter. Consignees were informed that they many continue to use their UHI-4 until the offset adjustment is performed. However, as noted in the Instructions for Use (IFU), prepare a spare UHI-4 unit as a backup to ensure uninterrupted operation can be completed without complication in the case of a malfunction. Actions to take included to carefully read the notification letter and ensure all personnel are completely knowledgeable on the content of the notification, an Olympus field service representative will contact you to schedule a mutuall
Quantity3,087 units
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, El Salvador, Germany, Honduras, Hong Kong, India, Ireland, Japan, Korea, Malaysia, Mexico, Myanmar, New Zealand, Paraguay, Peru, Russia, Singapore, Sri Lanka,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K243527High Flow Insufflation Unit (UHI-4)
Olympus Medical Systems Corporation, Jan 13, 2025