Recall Z-2678-2026 ยท DYB Introducer, Catheter

Oscor Inc.

Abiomed REF: 0052-3025, Introducer Kit for Impella Hemostatic Peel Away Introducer System with Infusion Sideport, 14F 13cm and 25cm. heart pump and accessories.

DateJul 10, 2026 (initiated Jun 10, 2026)
StatusOpen, Classified
ReasonDue to manufacturing issues, there is a potential for the 14Fr and 23Fr introducer sheaths to leak from the sidearm, under the sheath cap, and along the hub score lines in the hub region.
Root causeDevice Design
ActionOn June 10, 2026, Oscor (an Integer subsidiary ) issued a Urgent Medical Device Recall notification to affected consignees. Oscor asked consignees to take the following actions: 1. Product removal is not required, and hospitals may continue to use existing inventory. 2. Physicians should continue to use appropriate clinical awareness, monitoring, and patient support as needed. 3. Consider a device exchange or using the repositioning sheath to minimize blood loss if potential leakage from the sidearm and under the sheath cap and along the hub score lines in the hub region is observed. 4.Review
Quantity355 units
DistributionDistribution in the US state of MA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122084INTRODUCER SET, MODEL ADELANTE-S SERIES
Oscor, Inc., Aug 29, 2012