ClearanceBrief
Recall Z-2658-2026 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Bard Access Systems, Inc.

CK000662 PowerPICC SOLO HF Catheter 5 Fr Triple-Lumen UDI-DI Code: 00801741129674 Catheter stylets provide internal reinforcement to aid in catheter placement.

DateJul 2, 2026 (initiated May 21, 2026)
StatusOpen, Classified
ReasonDue to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Root causeNonconforming Material/Component
ActionOn 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the dru
Quantity66
DistributionWorldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K091324SHERLOCK 3CG TIP POSITIONING SYSTEM, SHERLOCK 3CG TIP POSITIONING SYSTEM STYLET
C.R. Bard, Inc., Aug 7, 2009