ClearanceBrief
Recall Z-2655-2026 ยท PXI Thoracentesis Tray

Bard Access Systems, Inc.

OTP5000 THORA/PARA TRAY, 5FR, NON-VALVED UDI-DI Code: 0110885403284748 PIG1260T SAFE-T-CENTESIS TRAY 6FR X 16CM UDI-DI Code: 0110885403108365 TPT1000 THORACENTESIS/PARACENTESIS TRAY UDI-DI Code: 0110885403108488 The Acute Drainage devices are indicated for use in removing fluid from either the pleural or peritoneal spaces during an acute drainage / aspiration procedure.

DateJul 2, 2026 (initiated May 21, 2026)
StatusOpen, Classified
ReasonDue to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.
Root causeNonconforming Material/Component
ActionOn 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the dru
Quantity9,500
DistributionWorldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.