ClearanceBrief
Recall Z-2590-2026 ยท LRO General Surgery Tray

Medline Industries, LP

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. CATARACT FULL BODY DYNJ85743 MAJOR VASCULAR PACK DYNJ81610C MINIMALLY INVASIVE PACK-LF DYNJ0843063L PEDIATRIC CATH ANGIO PACK DYNJ54755J PERIPHERAL ARTERIOGRAM RAD DYNJ61383C POSTERIOR SPINE PACK-LF DYNJ0843339O UPPER EXTREMITY DYNJ61907A VISCERAL ANGIOGRAM PACK-LF DYNJ0429121L

DateJun 25, 2026 (initiated Feb 27, 2026)
StatusOpen, Classified
ReasonFirm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.
Root causeNo Marketing Application
ActionThis entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affe
Quantity113,843 kits
DistributionUS Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213481Medline General Surgery Tray
Medline Industries, Inc., May 24, 2022