Recall Z-2553-2026 ยท LLZ System, Image Processing, Radiological

GE Medical Systems SCS

GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System

DateJun 24, 2026 (initiated May 8, 2026)
StatusOpen, Classified
ReasonGE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect tre
Root causeSoftware design
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 5/8/2026 by letter delivered via traceable means. The notice explained the issue, potential risk to patient, and requested the following: "Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use the device by following the instructions below: " Ensure that users verify the correct patient identity (including name, patient ID, and date of birth) and confirm that the selected study matches the intended patient and clinical context prior to reviewing or interpre
Quantity340 servers
DistributionWorldwide distribution - United States Nationwide and the countries of Algeria, Australia, Belarus, Belgium, Benin, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Korea (Republic Of), Lithuania, Mo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081985AW SERVER
Ge Medical Systems, LLC, Sep 25, 2008