Recall Z-2451-2026 ยท OWB Interventional Fluoroscopic X-Ray System

Medtronic Navigation, Inc.-Boxborough

Medtronic O-arm O2 Imaging System. Model Number: BI70002000.

DateJun 15, 2026 (initiated May 19, 2026)
StatusOpen, Classified
ReasonPotential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware.
Root causeUnder Investigation by firm
ActionMedtronic notified consignees on about 05/19/2026 via "URGENT: VOLUNTARY MEDICAL DEVICE CORRECTION" letter. Consignees were informed that affected systems may continue to be used prior to service by Medtronic to update the firmware. The issue is intermittent and may not present, however if it does it can be resolved by power cycling (rebooting) the system. A Medtronic Field Service Engineer will contact each facility to schedule time to update the firmware on each system to permanently resolve this issue. Consignees were instructed to complete and return the provided Customer Confirmation For
Quantity589 units
DistributionWorldwide distribution - US Nationwide and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Cyprus, Czech Republic, Denmark, Finland, France, Georgia, Germany, Greece, Guatemala, Hong Kong, India, Indonesia, Ir
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K240465O-arm O2 Imaging System
Medtronic Navigation, Inc., Jun 21, 2024