ClearanceBrief
Recall Z-2441-2026 ยท LRO General Surgery Tray

Medline Industries, LP

Medline convenience kits: ACL PACK DYNJ64161C ARTHROSCOPY PACK DYNJ47103D DYNJ67197G DYNJ85197A DYNJT4007 BASIC ACCESS PACK-LF DYNJ36698B BASIC BACK TRAY DYNJ89037 BASIC PEDI PACK-LF DYNJ0171446I DYNJ0171446J BASIC SPINE DYNJ910915A BREAST PACK DYNJ87587A BREAST PLASTIC PACK DYNJ84482C C SECTION DYNJ49172I CAROTID ENDARTERECTOMY PACK DYNJ48013L CENTRAL LINE PACK-LF DYNJ0220136S CERVICA

DateJun 12, 2026 (initiated Apr 27, 2026)
StatusOpen, Classified
ReasonKits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Root causeProcess control
ActionFirm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity49,654 kits total
DistributionUS Nationwide distribution. OUS distribution pending.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213481Medline General Surgery Tray
Medline Industries, Inc., May 24, 2022