Recall Z-2432-2026 ยท OHM Cesarean Section Tray

Medline Industries, LP

Medline convenience kits: C-SECTION KIT DYNJ904305J C-SECTION PACK DYNJ69533C

DateJun 12, 2026 (initiated Apr 27, 2026)
StatusOpen, Classified
ReasonKits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Root causeProcess control
ActionFirm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity49,654 kits total
DistributionUS Nationwide distribution. OUS distribution pending.
FDA recordOpen on fda.gov

510(k) clearances named in this recall