Recall Z-2913-2026 ยท PSY Whole Slide Imaging System

Ventana Medical Systems, Inc.

navify Digital Pathology (on-prem) v2.5, Material Number 10092343001 navify Digital Pathology (Cloud) v2.5. Material Number 09453733001 UDI-DI Code: 761333602095AQ In the United States, navify Digital Pathology software (previously known as uPath Enterprise Software) is a component of the Roche Digital Pathology Dx device cleared by the US FDA for Whole Slide Imaging purposes. Roche Digital Pa

DateAug 7, 2026 (initiated Apr 20, 2026)
StatusOpen, Classified
ReasonDue to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window
Root causeSoftware design
ActionOn 05/01/2026, the firm sent via UPS Ground an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that Roche eceived two complaints outside of the U.S. concerning the navify Digital Pathology (nDP) image management system regarding an issue where images may be mismatched or associated with incorrect specimen and slide labels within a given case. The discrepancy occurs exclusively within a subset of individual cases in which a specific nomenclature is used for specimen names and the case contains multiple specimens. Customers are instructed to: " Until navify Digital Pathology softw
Quantity190
DistributionWorldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, PR, SC, SD, TN, TX, UT, WI, and WV. The countries of Austria, Belgium, Czech Republic, France, Germany, Ireland, Italy, Netherlan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232879Roche Digital Pathology Dx (VENTANA DP 200)
Ventana Medical Systems, Inc., Jun 14, 2024
K242783Roche Digital Pathology Dx
Ventana Medical Systems, Inc., Dec 17, 2024