Recall Z-2430-2026 ยท MLT Pelvic Exam Kit
Medline Industries, LP
Medline convenience kit: SMSC-PELVISCOPY PACK-LF DYNJ0161619M
| Date | Jun 12, 2026 (initiated Apr 27, 2026) |
| Status | Open, Classified |
| Reason | Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application. |
| Root cause | Process control |
| Action | Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned. |
| Quantity | 49,654 kits total |
| Distribution | US Nationwide distribution. OUS distribution pending. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.