ClearanceBrief
Recall Z-2419-2026 ยท OFF Central Venous Catheter Tray

Medline Industries, LP

Medline convenience kits: INPATIENT CENTRAL LN DRSG KIT DYNDC3525 PICC/CVC/MIDLINE DRSG CHG KIT DT13191A PORT A CATH CDS982047S VEIN ABLATION PACK DYNJ62889C

DateJun 12, 2026 (initiated Apr 27, 2026)
StatusOpen, Classified
ReasonKits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Root causeProcess control
ActionFirm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity49,654 kits total
DistributionUS Nationwide distribution. OUS distribution pending.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.