ClearanceBrief
Recall Z-2503-2026 ยท OHD Gynecological Laparoscopic Kit

Medline Industries, LP

MEDLINE Medical Procedure Kits labeled as BASIC GYN, CW LAP GYN, D&C HYSTEROSCOPY, CW LAP GYN, DEACONESS PELVIC LAP PACK, FLOYD LAP GYN, GYN LAP PACK, GYN LAPAROSCOPY, GYN MINOR PACK, etc. (see the recall documents for a full list of affected products)

DateJun 17, 2026 (initiated May 4, 2026)
StatusOpen, Classified
ReasonDuring an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Root causeNonconforming Material/Component
ActionMedline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributor
Quantity44656 units
DistributionWorldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.