Recall Z-2480-2026 ยท OHR Catheter Care Tray
Medline Industries, LP
MEDLINE Urethral Catheterization Tray, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYNC1816, 15 Fr; 2. DYND10350, 15 Fr.
| Date | Jun 17, 2026 (initiated May 4, 2026) |
| Status | Open, Classified |
| Reason | During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification. |
| Root cause | Nonconforming Material/Component |
| Action | Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributor |
| Quantity | 8327756 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.