Recall Z-2398-2026 · MLM Enzyme Immunoassay, Tracrolimus

Siemens Healthcare Diagnostics, Inc.

Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay

DateJun 10, 2026 (initiated May 7, 2026)
StatusOpen, Classified
ReasonA potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
Root causeUnder Investigation by firm
ActionSiemens Healthineers issued an Urgent Medical Device Correction notice to its consignees on 5/12/2026 via FedEx. The notice described the issue, its impact to QC and patient results, and requested the following: "Customer Actions . Immediately discontinue use of and discard the affected kit lot numbers. . Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. . Review your inventory of these products to determine your laboratory s replacement needs and to provide information to Si
Quantity2878 units
DistributionWorldwide distribution - US Nationwide and the countries of Argentina, Austria, Brazil, Croatia, Germany, Hungary, India, Italy, Japan, Korea, Lithuania, Pakistan, Poland, Saudi Arabia, Spain, South Africa, Sri Lanka, Turkey, United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150168Dimension Tacrolimus Flex® Reagent Cartridge (TAC), Dimension Tacrolimus Calibrator (TAC CAL)
Siemens Healthcare Diagnostics, Nov 4, 2015