Recall Z-2289-2026 ยท OWB Interventional Fluoroscopic X-Ray System

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Philips Azurion 7B20/15, Model Numbers: 722068, 722226 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key

DateJun 1, 2026 (initiated Dec 7, 2023)
StatusOpen, Classified
ReasonDuring the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
Root causeSoftware design
ActionPhilips issued an IMPORTANT PRODUCT NOTICE to its consignees on 12/7/2023 via USPS mail. The notice explained the problem with the device and the circumstances under which if could occur and actions to be taken to minimize the effect of the problem. The notice further instructed the user to: - Keep this Important Product Notice with the documentation of the system until Philips corrects your system. - Circulate this notice to all users of the system so that they are aware of the issue. - Return the attached reply form to Philips to confirm that the users of the system have reviewed and und
Quantity156 units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov