Recall Z-2576-2026 ยท MWS Stabilizer, Heart
Medtronic Perfusion Systems
Octopus Evolution AS Tissue Stabilizer, Model TS2500
| Date | Jun 24, 2026 (initiated May 12, 2026) |
| Status | Open, Classified |
| Reason | During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design. |
| Root cause | Under Investigation by firm |
| Action | An URGENT: MEDICAL DEVICE RECALL notification letter dated 5/12/26 was sent to customers. Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for lot numbers listed in Attachment A. " If you have unused inventory from the listed lot numbers, immediately quarantine and return the products to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return and exchange or credit of this unit. Your Medtronic sales representative can assist you in the return of unit as necessary. |
| Quantity | 288 units |
| Distribution | Worldwide - US Nationwide distribution. |
| FDA record | Open on fda.gov |