Recall Z-2405-2026 ยท PNJ Leadless Pacemaker

Abbott Medical

Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software

DateJun 11, 2026 (initiated May 12, 2026)
StatusOpen, Classified
ReasonSoftware malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.
Root causeSoftware Design Change
ActionOn May 12 Abbott issued a Urgent Medical Device Recall notification via sales representative. Abbott ask consignees to take the following actions: Abbott has developed updated Merlin PCS 3650 programmer software (v28.9.1 rev 1 or higher) that resolves this behavior. Beginning with this notification, your Abbott Representative will upgrade programmer software to this software version . Abbott's service representative will proved a attestation to acknowledge receipt of the notification and software update. Please share this notification with others in your organization, as appropriate.
Quantity11182 programs
DistributionUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. OUS: Not provided yet
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.