Recall Z-2275-2026 ยท FRO Dressing, Wound, Drug

Oculus Technologies of Mexico, S.A. de C.V.

PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

DateJun 1, 2026 (initiated Apr 27, 2026)
StatusOpen, Classified
ReasonThe recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
Root causeComponent design/selection
ActionA MEDICAL DEVICE RECALL notification letter dated 4/27/26 was sent to customers. How to recognize that the device may fail The device may fail if the sprayer component shows internal corrosion, which can affect the appearance of the solution during use. Because the product is packaged in an amber glass bottle, it may be difficult to clearly see changes in the solution inside the container. Therefore, the issue is most likely to be identified during dispensing. Users may recognize this issue by observing a change in color of the solution when sprayed (for example, a light brown or tea-like app
Quantity5904 units
DistributionUS Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172851Microcyn Plus Antimicrobial Wound Care Solution
Sonoma Pharmaceuticals, Inc., Nov 15, 2017