ClearanceBrief
Recall Z-2472-2026 ยท HIF Insufflator, Laparoscopic

ConMed Corporation

ConMed AirSeal 12mm Access Port and Palm Grip Obturator with Bladeless Optical Tip, 150mm Length. Catalog Number: iAS12-150LPI. 6 units per pack. Disposable Endoscopic Trocar and Cannula; Carbon Dioxide Insufflator for Laparoscopy.

DateJun 17, 2026 (initiated May 14, 2026)
StatusOpen, Classified
ReasonPotential for overpressure alerts.
Root causeProcess change control
ActionConMed notified consignees on about 05/14/2026 via emailed and mailed letter titled URGENT: Medical Device Correction. Consignees were made aware of the issue and risk associated with the issue. They were instructed that a replacement AirSeal disposable should be available prior to procedures in order to minimize potential delays. Before initiating insufflation, the user should ensure the tubing is properly connected between the AirSeal tube set and Airseal Access Ports. If the connection cannot be securely tightened and the overpressure alert cannot be resolved, please replace the affected d
Quantity3,550 packs (21,300 eaches)
DistributionNationwide distribution. International distribution to Australia, Austria, Belgium, Bosnia and Herzegovina, Canada, CHILE, CHINA, COLOMBIA, Croatia, Denmark, DOMINICAN REPUBLIC, ECUADOR, Egypt, Estonia, Finland, France, Germany, Greece, HONG KONG, India, INDONESIA, Iraq, Ireland, Israel, Italy, Japa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143404SurgiQuest AirSeal iFS System
Surgiquest, Inc., Mar 20, 2015