Recall Z-2618-2026 ยท PHX Shoulder Prosthesis, Reverse Configuration

Medacta Usa Inc

Short Monobloc Stem Broaches, Size/REF: 7/04.01.10.0970, 8/04.01.10.0971, 9/04.01.10.0972, 10/04.01.10.0973, 11/04.01.10.0974, 12/04.01.10.0975, 13/04.01.10.0976, 14/04.01.10.0977, 15/04.01.10.0978, 16/04.01.10.0979 Part of Medacta Shoulder System

DateJun 29, 2026 (initiated Apr 30, 2026)
StatusOpen, Classified
ReasonSurgical instruments used during shoulder arthroplasty procedures does not progressively translate to replicate implant geometry, so the reaming guide and the stem reamer are positioned approximately 0.2 mm per size more laterally than intended, which may increase the risk of fracture in the medial region of the proximal humerus.
Root causeDevice Design
ActionOn 4/30/2026, field safety notices were emailed to customers who were asked to do the following: No pre-emptive interventions are recommended, however monitoring of the range of motion and stability should be evaluated during patient follow-up visits as per the standard of care. " Return all units of the lot numbers listed to firm. " Forward this notice to all people and organizations potentially affected. Firm Customer Service can be reached at 312-878-2381.
Quantity50
DistributionUS Nationwide distribution in the states of CO, AR, IN, PA, UT, FL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K250644MSS - Monobloc stem
Medacta International S.A., Feb 3, 2026