Recall Z-2224-2026 ยท FRN Pump, Infusion
Fresenius Kabi USA, LLC
Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
| Date | May 29, 2026 (initiated May 6, 2026) |
| Status | Open, Classified |
| Reason | Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without use |
| Root cause | Device Design |
| Action | On May 6, 2026 URGENT - Medical Device Field Correction letters were sent to customers. Required Actions for Healthcare Providers 1. If a unit has been dropped or severely jarred, remove it from service, even if no damage is visible. 2. Once you have removed the pump from service, or if you have questions about whether to do so after a drop, please contact Fresenius Kabi customer support at Ivenix_support@fresenius-kabi.com or (855) 354-6387. 3. Post enclosed Safety Tip Sheet, Proper Handling of the Ivenix Large Volume Pump (LVP) notice within your facilities to ensure that all ancillary staf |
| Quantity | 18,444 |
| Distribution | Domestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI; |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K183311 | Ivenix Infusion System (IIS) Ivenix, Inc., Jun 7, 2019 |