Recall Z-2224-2026 ยท FRN Pump, Infusion

Fresenius Kabi USA, LLC

Ivenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004

DateMay 29, 2026 (initiated May 6, 2026)
StatusOpen, Classified
ReasonFresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without use
Root causeDevice Design
ActionOn May 6, 2026 URGENT - Medical Device Field Correction letters were sent to customers. Required Actions for Healthcare Providers 1. If a unit has been dropped or severely jarred, remove it from service, even if no damage is visible. 2. Once you have removed the pump from service, or if you have questions about whether to do so after a drop, please contact Fresenius Kabi customer support at Ivenix_support@fresenius-kabi.com or (855) 354-6387. 3. Post enclosed Safety Tip Sheet, Proper Handling of the Ivenix Large Volume Pump (LVP) notice within your facilities to ensure that all ancillary staf
Quantity18,444
DistributionDomestic: CA, CO, FL, GA, ID, IL, MD, MI, MI , MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI, WI;
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019