Recall Z-2247-2026 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
GE Medical Systems Information Technologies Inc
GE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
| Date | May 28, 2026 (initiated Apr 24, 2026) |
| Status | Open, Classified |
| Reason | GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). |
| Root cause | Software design |
| Action | GE HealthCare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 4/24/2026 by letter using a traceable means. The notice explained the the safety issue pertaining to the persistent NO COMM or OFF NETWORK condition, and provided instructions for the safe use of the device should either issue arise. "Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your ApexPro CTS with Central Station for telemetry only or combination monitoring unless you observe a persistent (more than a few seconds) NO COMM or OFF NETWORK notificati |
| Quantity | 2187 units |
| Distribution | Worldwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K103142 | SPACELABS MULTI-PARAMETER MODULE Spacelabs Healthcare, Inc., Mar 3, 2011 |