Recall Z-2279-2026 ยท DQK Computer, Diagnostic, Programmable

GE Medical Systems Information Technologies Inc

CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3

DateJun 1, 2026 (initiated Apr 24, 2026)
StatusOpen, Classified
ReasonElectrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury.
Root causeProcess control
ActionFirm notified consignees starting on April 24, 2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION." Pending corrections from GE Healthcare, customers were instructed to perform a visual inspection of the printer assembly to verify proper positioning of the ESD brush. Customers may continue to use the device if the brush is properly positioned and aligned. However, if the brush appears misaligned or displaced, stop using the device and contact a GE HealthCare Service representative.
Quantity1835
DistributionWorldwide distribution - US Nationwide and the countries of Algeria, Australia, Bangladesh, Cambodia, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Ireland, Italy, Japan, South Korea, Lebanon, Libya, Lithuania, Malaysia, Myanmar, Netherlands, New Zealand,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231870CASE V7.0 Cardiac Testing System and CardioSoft V7.0 Cardiac Testing System
Ge Medical Systems Information Technologies, Inc., Dec 13, 2023