Recall Z-2223-2026 ยท FRN Pump, Infusion

Fresenius Kabi USA, LLC

LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.

DateJun 3, 2026 (initiated May 6, 2026)
StatusOpen, Classified
ReasonPotential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
Root causeSoftware design (manufacturing process)
ActionConsignees were notified on about 05/06/2026 via letter. Consignees were informed of the issue, risk associated, and instructed to keep the Ivenix pump plugged in at all times rather than rely on battery and if a "Battery Extremely Low" alarm occurs while on battery, immediately connect the pump to AC. For any transport of a patient on an affected pump, have a plan for AC access. For patients receiving critical or short half-life infusion therapies, ensure an RN accompanies the patient during transport. Consignees were also instructed to notify other users if applicable, notify customers if a
Quantity32 systems
DistributionUS Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183311Ivenix Infusion System (IIS)
Ivenix, Inc., Jun 7, 2019