Recall Z-2306-2026 ยท DXH Transmitters And Receivers, Electrocardiograph, Telephone

GE Medical Systems China Co., Ltd.

MAC 5 A4 Resting ECG Analysis System 855001-001; MAC 5 A5 Resting ECG Analysis System 8855002-001. Electrocardiograph

DateJun 3, 2026 (initiated Apr 21, 2026)
StatusOpen, Classified
ReasonSoftware in development inadvertently made available for customer use. Use of this software version can result in clinical information being associated with the incorrect patient. This could impact clinical decision-making.
Root causeSoftware Manufacturing/Software Deployment
ActionFirm notified consignees beginning April 21, 2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION". Customers were given instructions on how to check their software version. If a customer's software version begins with 3.00, the customer should not use the system until a GE HealthCare representative has updated the device with the latest licensed software version.
Quantity16
DistributionWorldwide - US Nationwide distribution in the states of Florida and Georgia. The countries of India, Italy, Republic of Korea, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221321MAC 5 A4/ MAC 5 A5/ MAC 5 Lite Resting ECG Analysis System
Ge Medical Systems Information Technologies, Inc., Oct 18, 2022