Recall Z-2298-2026 ยท BRY Cabinet, Table And Tray, Anesthesia
CareFusion 303, Inc.
BD Pyxis MedStation ES REF: 323 Medication cabinet
| Date | Jun 3, 2026 (initiated Apr 30, 2026) |
| Status | Open, Classified |
| Reason | Fingerprint scanner may overheat to a temperature to cause 1st degree burn. |
| Root cause | Component design/selection |
| Action | On April 30, 2026, CareFusion (subsidiary of Becton, Dickinson and Company) issued a Urgent Medical Device Recall Notification, via mail or Email. CareFusion ask consignees to take the following actions: 1. Dispensing cabinets can continue to be used as intended. 2. If you suspect that the Bio-ID Scanner is hot, please discontinue the use of Scanner and file a complaint. 3. Review and update facility policies and procedures related to situations where the Bio-ID Scanner is unavailable to mitigate delays in care. 4. Complete the attached Customer Response Form and return to the BD contact no |
| Quantity | 15,933 |
| Distribution | US: Nationwide OUS: AE, AU, BE, BH, BR, CA, CL, ES, GB, IT, KW, MA, MX, NZ, PT, QA, SA, SG, TH, TW |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.