ClearanceBrief
Recall Z-2718-2026 ยท KQT Evacuator, Gastro-Urology

Olympus Corporation of the Americas

Olympus Toomey Evacuator 30cc with Adaptor. Model Number: 188.

DateJul 10, 2026 (initiated May 5, 2026)
StatusOpen, Classified
ReasonSome sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
Root causeUnder Investigation by firm
ActionOlympus notified consignees on about 06/05/2026 via Courier, 2-day delivery. Consignees were requested to examine inventory for affected units. If affected units are on hand, replace any existing sterilization instructions for these products with those provided with the notification and add these instructions to any existing copies of the IFU. Consignees were also instructed to acknowledge receipt of the customer letter through the web portal. If affected units were further distributed, Olympus instructed consignees to identify those customers and forward the notification to them. An updated n
Quantity190 units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.